LASERS / PHACO
We provide our expertise in preventive and corrective maintenance of all lasers. Lasers in general when cared for properly and when scheduled maintenance parts are replaced at least semi-annually will bring many years of profit beyond your ROI.
A proper maintenance procedure includes replacement of the following:
- Clean and dust optics
- Wipe down lenses
- Replacement of DI filter
- Replacement of Particulate Filter
- Clean flow switch
- Flush cooling system
- Measure output
- Align optics
- Replace faulty mirrors
- Test safety filter
- Footswitch continuity test
- Replace blast shield as applicable